| Beamline Website |
http://www.maxlab.lu.se/beamline/max-ii/i911/i911-3.html |
| Owner/Operator |
Max-Lab at Lund University |
| Operational (general use) |
yes |
| %Time available (general use) |
100 |
| Beamline Schedule |
http://maxsun5.maxlab.lu.se/beamlines/bli911/schedules/archive.htm |
| Experiments |
macromolecular crystallography - MAD/SAD experiments |
| Source |
Multi-pole wiggler, period= 61 mm, 49 poles, Bmax = 3.5 T (Presently operated at 2.8 T)
|
| Ring Current(mA) |
200 ma |
| Average Beam Lifetime |
. |
| Spot Size |
0.3(h) x 0.2(v) mm2 (Measured) |
| Optics |
. |
| Flux |
1 x 1012 photons / sec in 0.3(h) x 0.3(v) mm2 (Calculated) |
| Flux Details |
. |
| Wavelength Min (Å) |
0.7 |
| Wavelength Max (Å) |
2.0 |
| Monochromator |
double crystal monochromator, water-cooled. |
| Crystal type |
Si (111) |
| Mirrors |
1st mirror: Rh-coated Si mirror, bent for vertical collimation 2nd mirror: Rh-coated toroidal Si mirror.
|
| Mirror Cutoff |
. |
| Special Capabilities |
. |
| Other Instrumentation |
RAntec fluorescence detector |
| Goniometer |
crystal logic goniostat |
| Pin Length |
.18 mm pins |
| Restrictions |
motorized goniometer head not in operation as yet |
| xtal-detector Distance |
. |
| Detector |
Mar 225 CCD |
| Detector Details |
. |
| 2theta range |
. |
| Cryo Capability |
Oxford instruments cryoject |
| Lab Facilities |
Olympus microscope with cold light source |
| Biohazard |
. |
| Software |
Mosflm, CCP4 |
| Hardware |
. |
| Backup Devices |
FTP, CD/DVD, firewire |
|
|