Primary Contact: Name | Jose Brandao-Neto |
Secondary Contact: Name | Alice Douangamath |
Owner/Operator | Diamond, Harwell Chilton Science Campus, South Oxfordshire, UK |
Status | Operational |
Status Details | I04-1 is currently operating in optimisation mode. |
%Time Available (general use) | 25 |
Experiments | SAD, FLUORESCENCE |
Experiments Details | I04-1 will be dedicated to high-throughput data collection for rapid structure solution using MR method. |
Special Capabilities | Absorption Edges |
Special Capabilities Details | Single anomalous diffraction (SAD) experiments will be possible at the Bromine K edge. |
Spot Size (mm): Height | 0.05 |
Spot Size (mm): Width | 0.05 |
Monochromator | single bounce |
Crystal Type | Si(111) |
Mirrors | toroidal |
Flux(PHOTONS/SEC) | 3.5 x 10^11 |
Flux Details | @ 13530 eV at sample |
Wavelength Range (Å) | Fixed: 0.916 |
Energy Range (keV) | Fixed: 13.53 |
Goniometer | Single axis, extremely accurate air bearing goniometer (a possible addition of a mini-kappa is being explored) |
Sample Handling: Robotics | (CATS) Cryogenic Auto Transfer System |
Robotics: More Info | SPINE pin, EMBL/ESRF and Rigaku magazines compatibility |
1. Detector: Type | CCD |
1. Detector: Model | MARmosaic 300 |
1. Detector: Activated for Service | 2010-04-10 |
1. Detector: Removed from Service | 2010-10-31 |
2. Detector: Type | PIXEL |
2. Detector: Model | DECTRIS PILATUS 2M |
2. Detector: Details | Pilatus 2M |
2. Detector: Activated for Service | 2010-11-20 |
2. Detector: Removed from Service | |
Cryo Capability | A cryocooling device to keep the sample at 100K |