APS Beamlines:
5ID-B
,
8-BM
,
14-BM-C
,
14-BM-D
,
14-ID-B
,
17-BM
,
17-ID
,
18-ID
,
19-BM
,
19-ID
,
21-ID-F
,
21-ID-G
,
22-BM
,
22-ID
,
23-BM
,
23-ID-B
,
23-ID-D
,
24-ID-C
,
24-ID-E
,
31-ID
|
| Beamline Website |
http://www.imca.aps.anl.gov/ |
| Owner/Operator |
IMCA-CAT |
| Operational (general use) |
yes |
| %Time available (general use) |
25% |
| Beamline Schedule |
http://www.imca.aps.anl.gov |
| Experiments |
monochromatic & MAD |
| Source |
undulator |
| Ring Current(mA) |
100 |
| Average Beam Lifetime |
During top-up operations, lifetime is infinite. |
| Spot Size |
0.5 mm x 0.5 mm maximum, 0.05mm x 0.05 mm minimum (unfocused; defined by slits) |
| Optics |
Si(111) double-crystal monochromator |
| Flux |
2.3*10E(11) photons/s/(200 micron aperture) |
| Flux Details |
. |
| Wavelength Min (Å) |
2.0 |
| Wavelength Max (Å) |
0.7 |
| Monochromator |
? |
| Crystal type |
? |
| Mirrors |
? |
| Mirror Cutoff |
? |
| Special Capabilities |
? |
| Other Instrumentation |
? |
| Goniometer |
Equipment in the experiment end station includes CrystalLogic goniostat with omega and theta rotations, Rigaku/MSC ACTOR robot for automated crystal mounting, positioning, and retrieval, and Rontec XFlash fluorescence detector.
|
| Pin Length |
18 mm |
| Restrictions |
. |
| xtal-detector Distance |
85 mm to 585 mm |
| Detector |
ADSC Q210 CCD detector |
| Detector Details |
51-micron pixels in a 4096 grid |
| 2theta range |
vertical movement of 0 degrees to 35 degrees |
| Cryo Capability |
Oxford Instruments CryoJet |
| Lab Facilities |
Biochemistry laboratory and cold room for sample preparation. |
| Biohazard |
BSL1. Heavy atom soaks are permitted. |
| Software |
Data collection: JDirector.
Data processing: HKL-2000, d*TREK, MOSFLM, XDS.
Beamline control: MX.
|
| Hardware |
Linux workstations |
| Backup Devices |
CD-R, DVD-R, external Hi-speed USB, and FireWire (IEEE-1394a) hard drives, user's laptop computer via Gigabit Ethernet
|
|
|
|
Update this site - synchrotron administrators only
|