APS Beamlines:
5ID-B
,
8-BM
,
14-BM-C
,
14-BM-D
,
14-ID-B
,
17-BM
,
17-ID
,
18-ID
,
19-BM
,
19-ID
,
21-ID-F
,
21-ID-G
,
22-BM
,
22-ID
,
23-BM
,
23-ID-B
,
23-ID-D
,
24-ID-C
,
24-ID-E
,
31-ID
|
| Beamline Website |
http://cars.uchicago.edu/biocars/pages/bmc.html |
| Owner/Operator |
BioCARS |
| Operational (general use) |
yes |
| %Time available (general use) |
100 |
| Beamline Schedule |
http://biocars.org |
| Experiments |
Macromolecular Crystallography, Monochromatic (0.9A) |
| Source |
bending magnet |
| Ring Current(mA) |
100 |
| Average Beam Lifetime |
infinite in top-up mode |
| Spot Size |
0.13mm H x 0.34mm V |
| Optics |
bent conical Si-mirror (Rh coating); bent Ge(111) monochromator |
| Flux |
6X10^11 photons/sec @ 13.78KeV |
| Flux Details |
. |
| Wavelength Min (Å) |
0.8 |
| Wavelength Max (Å) |
1.6 |
| Monochromator |
bent Ge(111) |
| Crystal type |
Ge(111) |
| Mirrors |
bent conical Si, Rh coating |
| Mirror Cutoff |
. |
| Special Capabilities |
. |
| Other Instrumentation |
on-line micro-spectrophotometer |
| Goniometer |
kappa goniometer w/ automated crystal centering |
| Pin Length |
15-35 mm |
| Restrictions |
. |
| xtal-detector Distance |
100 to 1000 mm adjustable |
| Detector |
ADSC Quantum 315 |
| Detector Details |
MAR-345 available on request |
| 2theta range |
. |
| Cryo Capability |
Cryojet (Oxford Instruments) |
| Lab Facilities |
laser laboratory (ps and ns lasers), cold room, biosafety cabinet |
| Biohazard |
support for BSL-2 and BSL-3 experiments |
| Software |
Beamline control:EPICS; Data acquisition: BioCARS-developed GUI; Data processing: HKL2000, Denzo/HKL, Mosflm, EPMR, Phaser, Solve/Resolve, CCP4, SHELX, O, XtalView, LaueView, Precognition
|
| Hardware |
. |
| Backup Devices |
user-provided firewire and USB drives |
|
|
|
Update this site - synchrotron administrators only
|