| Beamline Website |
http://www-als.lbl.gov/als/techspecs/bl5.0.3.html |
| Owner/Operator |
GNF and Takeda |
| Operational (general use) |
yes |
| %Time available (general use) |
25 |
| Beamline Schedule |
http://bcsb.als.lbl.gov/ |
| Experiments |
Monochromatic (1A fixed) |
| Source |
56 poles 11.4 cm period length wiggler |
| Ring Current(mA) |
400 |
| Average Beam Lifetime |
8 Hours |
| Spot Size |
100 micron spot size with 2.7mrad (h) x 0.4mrad (v) maximum divergence |
| Optics |
? |
| Flux |
1.7e11 photons/sec |
| Flux Details |
measured flux through 100um diameter collimator at 400mA ring current and 1.5mrad divergence
|
| Wavelength Min (Å) |
1.0 |
| Wavelength Max (Å) |
1.0 |
| Monochromator |
Asymmetric curved crystal |
| Crystal type |
Si(220) |
| Mirrors |
Rh/Pt coated |
| Mirror Cutoff |
? |
| Special Capabilities |
? |
| Other Instrumentation |
? |
| Goniometer |
Single axis; air bearing. |
| Pin Length |
(16-28) mm |
| Restrictions |
no shorter than 16mm pins only |
| xtal-detector Distance |
90 - 1070mm. |
| Detector |
ADSC Quantum 315R;CCD area detector |
| Detector Details |
3x3 CCD array, 315x315mm, 6144x6144 pixels, pixel size 51x51um |
| 2theta range |
-5 to 30 degrees |
| Cryo Capability |
Oxford Instruments Cryojet; ~95K to room temperature |
| Lab Facilities |
Biochemical lab (comprehensive) with all equipment necessary for structural studies. Cold room available. Preferable to reserve in advance.
|
| Biohazard |
Contact in advance. |
| Software |
ADSC, HKL2000, CCP4, MosFlm, DENZO, XDS;
processing: CCP4, PHENIX, CNS, SOLVE/RESOLVE, etc.
|
| Hardware |
Crystal mounting robot. Four Dual Processor Linux. 2Tbyte linux raid storage |
| Backup Devices |
PC,MAC,Linux FireWire drives, 4 & 8mm tape, removable SCSI Hard disk. |
|
|
|
Update this site - synchrotron administrators only
|